- 'MR Conditional' ≠ 'MR Safe in all conditions'.
- Conditions include 1.5 T and 3 T only, and limits on gradient and SAR.1
- Always tell the radiology team you have dental implants.
The conditions, as published by Straumann
Straumann's MRI safety information states that patients with implants from the Straumann® Dental Implant System can be scanned safely under specific conditions (MR Conditional): 1.5 T and 3 T only, with limits on spatial gradient and SAR. 12
'MR Conditional' is not the same as 'safe under all conditions'. The radiology team must check the conditions.
| Parameter | Condition |
|---|---|
| Static magnetic field | 1.5 Tesla and 3 Tesla only1 |
| Maximum spatial gradient | 4,000 gauss/cm (40 T/m)1 |
| Maximum SAR | 2 W/kg whole body, 3.2 W/kg head, 15 min per sequence, Normal Operating Mode1 |
| Maximum heating | 4.9 °C after 15 minutes of continuous scanning1 |
| Image artifact | About 10 mm from the implant (gradient echo, 3 T)1 |
These conditions are for the radiology team. This page is not personal medical advice.
What to tell the radiographer
- That you have dental implants, how many, and where.
- That they are Straumann implants and, if you know it, the system.
- Any other metal parts, devices or implants you have.

Would you like a dentist to assess whether this treatment suits you?
Request an assessmentLast updated 28 Sep 2026 · How we review
Sources
- Manufacturer documentationMRI Safety Information, document 702071/H/07 (05/23) — Institut Straumann AG, 2023-05. Accessed 28 Sep 2026.
- Regulatory recordFDA 510(k) K190662 — MRI Compatibility for Existing Straumann Dental Implant Systems — U.S. Food and Drug Administration, 2019-06-17. Accessed 28 Sep 2026.
Educational information only. The choice of implant system depends on your anatomy, restorative plan, bone quality and the clinician's assessment. This website does not recommend a specific implant for you.